Instructions for use Babylog 8000 plus SW 5.n
75
Operation
Performing medication nebulization
with an Aeroneb Pro nebulizer
Before nebulization with an Aeroneb Pro
1
Observe the instructions for use of the Aeroneb
Pro nebulizer.
2
Observe chapter ''Preparing the ventilator'' on
page 42.
3
Observe the safety information on medication
nebulization, see page 70.
4
Do not switch on the nebulization function on
Babylog 8000 as the Aeroneb Pro nebulizer
does not require a nebulizer flow from
Babylog 8000.
5
Remove the flow sensor from the breathing
circuit, see page 71.
After nebulization with an Aeroneb Pro
1
Install the flow sensor, see page 71.
2
Switch flow monitoring on, see page 71.
3
If a bacterial filter is used to protect the
expiratory valve, replace or remove the
bacterial filter.
4
If no bacterial filter is used, attach a
reprocessed expiratory valve, see page 42.