Instructions for use Babylog 8000 plus SW 5.n
47
Assembly and preparation
Replacing the flow sensor insert
The insert of the flow sensor must be replaced
when the following alarm message is displayed:
Flow measurement disturbed
Measurement switched off
1
Remove the plug (A) of the flow sensor cable
from the flow sensor.
2
Gently press the knobs (B) on both sides while
pulling the insert out of the flow sensor
housing.
3
Push the new insert (C) in until it engages.
Ensure that the two markings are aligned.
4
Connect the plug (A) of the flow sensor cable to
the flow sensor.
5
Calibrate the flow sensor, see page 95.
Connecting the test lung
The test lung consists of the following
components:
– Bellows:
Compliance: 0.5 mL/mbar (0.5 mL/cmH
2
O)
– Endotracheal tube CH 12:
Length: approx. 165 mm (6.5 in)
– Connector
Connecting the test lung
Insert the test lung in the patient connection
port of the breathing circuit.
20919
A
B
C
20912