
Maintenance
124
Instructions for use Babylog 8000 plus SW 5.n
Preventive maintenance
Table of preventive maintenance intervals
WARNING
Risk of faulty components
Device failure is possible due to wear or
material fatigue of the components.
To maintain the function of all components,
this device must be inspected and serviced at
the intervals specified by the manufacturer.
WARNING
Risk of electric shock
Before performing any service work,
disconnect all electrical connections and gas
connections from power and gas supplies.
CAUTION
Risk of personal injury
If the glass on the LC screen is damaged, a
chemical liquid may escape.
– Avoid contact with the body.
– Immediately clean affected skin areas with
soap.
Component
Interval
Measure
Personnel responsible
O
2
sensor
When the O
2
sensor can
no longer be calibrated
Replace, see page 42
Users
Cooling air filter
Every 4 weeks
Clean, see page 126
Users
Every 12 months
Replace
O-ring for expiratory
valve
Every 12 months
Replace
Experts
Diaphragm of the expira-
tory valve
Every 12 months
Replace
Experts
Lip seal for pneumatic
part
Every 2 years
Replace
Service personnel
Battery for mains power
supply failure alarm
Every 2 years
Replace
Experts
Flat seal on entry block
of gas connection
Every 2 years
Replace
Experts
Filter on entry block of
gas connection
Every 2 years
Replace
Experts
Time keeper RAM
Every 4 years
Replace
Experts
Pressure reducers
Every 6 years
Basic overhaul of pres-
sure reducers
Experts