Instructions for use Babylog 8000 plus SW 5.n
89
Trends
Displaying measured volume values for high-
frequency ventilation (
HFV
)
Condition: High-frequency ventilation option must
be available.
1
On the main screen, press the
Meas
key.
2
Press the
HFVol
key.
The following measured values are displayed
(example):
Displaying all set values and measured
values
Displaying all set values
On the main screen, press the
Values
key.
The
Set1
dialog window is displayed.
Displaying additional set values:
Press the
Set2
key.
Displaying all measured values
1
On the main screen, press the
Values
key.
2
Press the
Meas1
key.
Displaying additional measured values:
Press the
Meas2
key.
21083
MVim
Inspiratory measured minute volume
from mandatory ventilation [L/min]
DCO
2
Gas transport coefficient [mL
2
/s]
DCO
2
=
VThf
2
x
f
f
: Frequency of high-frequency pulses
[Hz]
VTim
Inspiratory measured tidal volume of
mandatory ventilation [mL]
VThf
Inspiratory measured tidal volume of
high-frequency ventilation [mL]
21084
21085
21086
21087