Instructions for use Babylog 8000 plus SW 5.n
149
Technical data
Radiated radio fre-
quency disturbance
(IEC 61000-4-3)
80 MHz to 2.5 GHz:
10 V/m
10 V/m
Recommended minimum
distance to portable and
mobile radio frequency
transmitters with transmis-
sion power P
EIRP
to the
medical device including its
lines:
1)
Conducted radio fre-
quency disturbance
(IEC 61000-4-6)
150 kHz to 80 MHz:
10 V inside ISM
bands
2)
10 V
Recommended minimum
distance from portable and
mobile radio frequency
transmitters with transmis-
sion power P
EIRP
to the
medical device including its
lines:
1)
150 kHz to 80 MHz:
3 V outside ISM
bands
2)
3 V
1) For P
EIRP
, insert the highest possible "equivalent isotropic radiated power" of the adjacent radio frequency transmitter. In
the vicinity of equipment marked with the symbol
, interference can occur. Field strengths from fixed, portable, or mobile
radio frequency transmitters at the location of the medical device should be less than 3 V/m in the frequency range from
150 kHz to 2.5 GHz and less than 1 V/m above 2.5 GHz.
2) ISM bands in this frequency range are: 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to
27.283 MHz; 40.66 MHz to 40.70 MHz.
Immunity against
IEC 60601-1-2 test
level
Compliance level
(medical device)
Electromagnetic environ-
ment