Assembly and preparation
44
Instructions for use Babylog 8000 plus SW 5.n
Attaching the bacterial filter
Conditions:
– Kit 8410230 must be used.
– The instructions for use for the bacterial filter
must be observed.
1
Attach the breathing hose (A) 0.25 m (9.8 in)
onto the inspiratory port.
2
Insert the adapter Ø15/Ø22 (B) into the
breathing hose.
3
Attach the bacterial filter (C) onto the adapter.
4
Insert the catheter connector size II (D) into the
bacterial filter.
5
Attach the breathing hoses.
Attaching the breathing hoses
Conditions:
– The breathing circuit in use must be suitable for
the individual patient.
– When no incubator is being used, a hinged arm
with clamp must be used.
CAUTION
Risk of contaminating the device
In the event of device failure, expiratory air may
enter the device via the inspiratory line.
Use an inspiratory bacterial filter.
20909
A
B
C
D
WARNING
Risk of electric shock and fire
The use of antistatic or conductive breathing
hoses increases the risk of electric shock to
the patient and of fire in an oxygen-enriched
environment.
Do not use antistatic or conductive breathing
hoses.
CAUTION
Humidification is ineffective if the inspiratory and
expiratory ports are reversed.
Attach the breathing hoses correctly.