Instructions for use Babylog 8000 plus SW 5.n
135
Technical data
Waveform display (on screen)
Range
–10 to 100 mbar (–10 to 100 cmH
2
O)
Resolution
–2.5 to 25 mbar (–2.5 to 25 cmH
2
O): 0.5 mbar
(0.5 cmH
2
O)
–5 to 50 mbar (–5 to 50 cmH
2
O): 1 mbar
(1 cmH
2
O)
–10 to 100 mbar (–10 to 100 cmH
2
O): 2 mbar
(2 cmH
2
O)
Flow and volume measurement
Sensor
Hot wire anemometer
Neonatal flow sensor Y-piece (8410185)
Dead space
1.7 mL
Resistance
≤
12 mbar (
≤
12 cmH
2
O) at 30 L/min
Neonatal flow sensor ISO 15 (8411130)
Dead space
0.9 mL
Resistance
≤
11 mbar (
≤
11 cmH
2
O) at 30 L/min
Flow
Range
0.2 to 30 L/min
Tidal volume
VT
Range
0 to 999 mL
Resolution
0.1 mL
Accuracy
≤
5 mL: ±0.5 mL
>5 mL: ±10 % of measured value
Minute volume
MV
Range
0 to 30 L/min
Resolution
0 to 0.99 L/min: 0.01 L/min
1 to 9.9 L/min: 0.1 L/min
10 to 30 L/min: 1 L/min
Accuracy
(Sensor in standard breathing circuit,
endotracheal tube >3 mm (0.12 in))
≤
2 L/min: ±0.2 L/min (
NTPD
)
>2 to 30 L/min: ±10 % (
NTPD
)
Leakage
Leak
Range
0 to 100 %
Resolution
1 %
Spontaneous portion of minute volume
spont
Range
0 to 100 %
Resolution
1 %