Instructions for use Babylog 8000 plus SW 5.n
53
Assembly and preparation
Transport within the hospital
Transport refers to any movement of the medical
device without the patient that does not solely
serve to position the medical device.
Increasing stability
1
Set the hinged arm to minimum extension.
2
Empty the water container of the breathing gas
humidifier.
3
Do not attach any additional parts to the lateral
standard rails.
4
Remove the monitor if there is one.
5
Grasp the device firmly by the lateral standard
rails and push the device in longitudinal
direction.