Instructions for use Babylog 8000 plus SW 5.n
103
Configuration
Setting the default values during operation
Press the
key.
The default values are used.
If a measured value exceeds the limits of the
scale, the voltage is limited to the maximum value
on the scale.
Configuring the impulse output
The following signals are available:
– Triggered mandatory breath
– Mandatory breath
– Alarm
Default value: Mandatory breath
1
Open the dialog window by pressing the
Cal.
Config.
>
Config
>
Com
keys.
2
Press the
Param
key repeatedly until
is
highlighted.
3
Use the
or
key to select the signal.
DCO
2
0 ... 200 mL
2
/s
→
0 ... 10 V
0 ... 50 mL
2
/s
→
0 ... 10 V
Continuous flow (set value)
0 ... 125 L/min
→
0 ... 10 V
Leakage rate
0 ... 100 %
→
0 ... 10 V
Spontaneous portion of minute vol-
ume
0 ... 100 %
→
0 ... 10 V
Factory-set default values:
Analog1
: Flow
–20 ... 20 L/min
Analog2
: Airway pressure
–10 ... 90 mbar (–10 ... 90 cmH
2
O)
Ventilation parameters
Scale range
Measuring signal
range
21
153
/P
21
1
5
4