Instructions for use Babylog 8000 plus SW 5.n
59
Getting started
2
Disconnect the probe of the O
2
compressed
gas hose from the wall terminal unit of the
central gas supply system.
The visual alarm signal flashes red, the acoustic
alarm signal sounds, and the alarm message
O
2
pressure low
is displayed.
3
Plug the probe into the wall terminal unit of the
central gas supply system.
The visual alarm signal goes out, the acoustic
alarm signal is silenced, and the alarm message is
no longer displayed.
Checking the compressed air alarm:
1
Disconnect the probe of the compressed air
gas hose from the wall terminal unit of the
central gas supply system.
The visual alarm signal flashes red, the acoustic
alarm signal sounds, and the alarm message
Medical air low
is displayed.
2
Plug the probe into the wall terminal unit of the
central gas supply system.
The visual alarm signal goes out, the acoustic
alarm signal is silenced, and the alarm message is
no longer displayed.
Checking the
IPPV
ventilation mode
1
Set the following ventilation parameters:
2
On the main screen, press the
Meas
key.
3
Press the
Paw
key.
The displayed values must be within the following
ranges:
Checking the PEEP
1
Set the
PEEP/CPAP
setting knob to 10.
2
Press the
OK
key.
3
On the main screen, press the
Meas
key.
4
Press the
Paw
key.
The displayed values must be within the following
ranges:
Checking the alarm limits
Checking apnea monitoring:
Set the
CPAP
ventilation mode.
After a maximum of 30 seconds, the visual alarm
signal flashes red, the acoustic alarm signal
sounds, and one of the following alarm messages
is displayed:
Checking the alarm limits for airway pressure:
1
Set the
IPPV/IMV
ventilation mode.
Parameter
Setting
Ventilation mode
IPPV/IMV
O
2
-Vol.%
21
10
T
I
0.4
T
E
0.6
Pinsp
20
PEEP/CPAP
0
Insp. Flow
Measured
value
Permissible range
Peak
18 to 22 mbar (18 to 22 cmH
2
O)
Mean
6 to 10 mbar (6 to 10 cmH
2
O)
PEEP
–1.5 to 1.5 mbar
(–1.5 to 1.5 cmH
2
O)
Measured
value
Permissible range
Peak
18 to 22 mbar (18 to 22 cmH
2
O)
Mean
12 to 16 mbar (12 to 16 cmH
2
O)
PEEP
8 to 12 mbar (8 to 12 cmH
2
O)
–
Apnoea
–
MV low