Technical data
138
Instructions for use Babylog 8000 plus SW 5.n
Monitoring functions
Sound pressure level L
PA
of alarm signals mea-
sured in accordance with IEC 60601-1-8:
Defined operating position: Front of the device at a
distance of 1 m (39 in) and a height of 1.5 m (59 in).
Sound pressure
Range for high-priority alarms according to
volume setting
63 dB (A) to 83 dB (A)
Range for low- and medium-priority alarms
according to volume setting
63 dB (A) to 82 dB (A)
Power supply failure alarm
53 dB (A)
Expiratory minute volume
MV
Upper alarm limit alarm
If the upper alarm limit has been exceeded
Range
0.10 to 0.99 L/min in steps of 0.01 L/min
1.0 to 9.9 L/min in steps of 0.1 L/min
10 to 15 L/min in steps of 1 L/min
Lower alarm limit alarm
If the lower alarm limit has been exceeded
Range
0 to 0.99 L/min in steps of 0.01 L/min
1.0 to 9.9 L/min in steps of 0.1 L/min
10 to 14 L/min in steps of 1 L/min
Alarm suppression
For the duration of the set alarm delay time
Apnea alarm time
Apnoea time
Alarm
If no breathing activity is detected
Range
5 to 20 s in steps of 1 s
>20 s: monitoring is deactivated
Respiratory rate (panting breathing)
Panting
Alarm
If the alarm limit has been exceeded
Range
20 to 200 bpm
<20 bpm: monitoring is deactivated
Alarm delay time
Alarm delay
Delays the following alarms:
–
MV high
–
VT low
Check settings!
Range
0 to 30 s