MAR-000276_v.1.3
Chapter:
Appendices
Page 46 of 48
13
Z bed probes are too inconsistent
Check that the bed is
correctly fitted
14
The memory buffer on the motor control
board is unexpectedly full
Restart, contact
support@dolomite-
microfluidics.com
problem persists
15
The motor control board temperature is
too high
Restart, contact
support@dolomite-
microfluidics.com
problem persists
30
Communications with the extruder board
have been lost
Restart, contact
support@dolomite-
microfluidics.com
problem persists
31
Extruder board rejected packet with
unexpected size
Restart, contact
support@dolomite-
microfluidics.com
problem persists
32
Extruder board rejected packet with
incorrect checksum after retries
Restart, contact
support@dolomite-
microfluidics.com
problem persists
33
Extruder board has rejected the unknown
command
Restart, contact
support@dolomite-
microfluidics.com
problem persists
34
Extruder board rejected read of an invalid
address
Restart, contact
support@dolomite-
microfluidics.com
problem persists
35
Extruder board rejected write to an invalid
address
Restart, contact
support@dolomite-
microfluidics.com
problem persists
36
Print head board is in bootloader mode
and needs to be updated
Restart, contact
support@dolomite-
microfluidics.com
problem persists
100
Nozzle has not been detected on
extruder
Check that the nozzle is
correctly fitted, see Section
3.3, figures 10 and 11
101
Inductor has gone into error
Restart, contact Support if
this problem persists
102
Inductor has read an invalid temperature Restart, contact
support@dolomite-
microfluidics.com
problem persists
103
Inductor has gone over current
Restart, contact