3.4 PRESSURE MENU
The pressure is controlled from this menu. Use SELECT to select the line to be changed. When selected
the line changes to blue (edit mode). Use DEFAULT or UP and DOWN arrows to change the setpoint value.
Pressing ENTER will save the new setpoint.
P1 Inlet setpoint is the pressure value that the regulator is setting on the inlet side of the Pressure
Myograph. Pressure range is 0-250mmHg.
P2 Outlet setpoint is the pressure value that the regulator is setting on the outlet side of the Pressure
Myograph. Pressure range is 0-250mmHg.
Pressure gradient set the speed at which the regulator modifies the pressure. The pressure change range
is 1-10mmHg/sec.
IMPORTANT:
P1 inlet setpoint should be higher than p2 outlet setpoint to create flow. If p1 inlet
setpoint is lower than P2 OUTLET SETPOINT, IT WILL PREVENT BACKFLOW IN THE SYSTEM.