3.7.2.3 Set-up the pressure calibration on the Pressure
Interface for the 112PP System
Step 1 - Press NEXT STEP when the system is ready.
Step 2 - Apply a pressure of 50mmHg on the Pressure
Myograph using the Big Ben Pressure Manometer. When the
4 blue lines of relative values stabilize, go to NEXT STEP.
The relative values should be between 3300 to 3900.
Step 3 - Apply a pressure of 125mmHg to the Pressure
Myograph using the Big Ben Pressure Manometer. When the
4 blue lines of relative values stabilize, go to NEXT STEP.
The relative values should be between 3900 to 4500.
Step 4 - Apply a pressure of 200mmHg to the Pressure
Myograph using the Big Ben Pressure Manometer. When the
4 blue lines of relative values stabilize, go to NEXT STEP.
The relative values should be between 4600 to 5200.