52
3.7.7. PULSE RATE (ONLY FOR PULSATILE 112PP SYSTEM)
The Select Pulse Rate Menu is only for users that have the
Pulsatile Pressure 112PP chamber/system. In the Pulse Rate
menu, the number of pulses per minute can be set between
50 and 600 Pulses per minute. By selecting 400 Pulse/min, the
Pressure Interface will pulse between the P1 and P2 pressure
400 times a minute.
The Duty Pulse High is the percentage of the time the Pressure
should be at the P1 pressure. With a setting of e.g. 30%, the
Pressure Interface will be at the P1 pressure 30% of the given
time and 70% of the time at the P2 pressure. Changing the
Duty Pulse High value will change the pulsing profile if wanted.
PULSATION: Pulsation is turned ON or OFF
All the above settings can be controlled in the MyoVIEW 4.0
software.
3.7.8. CLEANING MENU (ONLY PULSATILE 112PP)
The Cleaning Menu is mandatory to use when the system is
filled/purged with running buffer or cleaned. In the Cleaning
Menu the user is able to monitor all four pressure transducers
live in the 112PP Pulsatile Pressure System. This is very useful
to make sure that manual excessive pressure is not applied
to the system at any time. Applied pressure above 250mmHg
will break the Pressure Transducers on the 112PP system.
Furthermore, the valve switching between P1 and P2 pressure
will be active making the cleaning of the system possible.