5. Lift up the source needle to
remove it from the needle
block (see
then pull the needle through
the rear of the needle
assembly.
Figure 5-24. Removing the Source Needle from the Needle Block
6. Use a 1/4-inch open-end wrench to disconnect the fitting on the side of the
static valve (see
7. Remove the source needle and tubing from the ASE 150.
Figure 5-25. Disconnecting the Source Needle Fitting from the Static Valve
Source
Needle
Disconnect
this fitting