ASE 150 Operator’s Manual
28
Doc. 065207-02 9/08
H
2
SO
4
0.1 M
Yes
Yes
Yes
KOH 8 M
No
No
N/A
KOH 10%
No
Yes
N/A
KOH 0.1 M
Yes
Yes
N/A
HNO
3
8 M
No
No
Yes
HNO
3
10%
Yes
Yes
Yes
HNO
3
0.1 M
Yes
Yes
Yes
NH
4
OH 8 M
No
No
N/A
NH
4
OH < 5%
Yes
Yes
N/A
NH
4
OH 0.1 M
Yes
Yes
N/A
Trifluoroacetic acid
< 5%
Yes
Yes
Yes
Acetic acid < 5%
Yes
Yes
Yes
Phosphoric acid
<5%
Yes
Yes
Yes
If acids, bases, salts, or buffers are used as extraction solvents, rinse
the system with 100% polar organic solvent (for example, acetone or
methanol) or distilled water before turning off the power.
Acid or Base Con-
centration
Okay to pump from
solvent reservoir?
(Okay to use SST
or Zr cells)
Add to sample and
transfer to Zr cell?
(Do not use SST
cells)
Add to sample
matrix with ASE
Prep CR and trans-
fer to Zr cell? (Do
not use SST cells)
Table 3-1. Guidelines for Use of Acids and Bases with the ASE 150 (Continued)
Summary of Contents for ASE 150
Page 32: ...ASE 150 Operator s Manual 24 Doc 065207 02 9 08...
Page 64: ...ASE 150 Operator s Manual 56 Doc 065207 02 9 08...
Page 76: ...ASE 150 Operator s Manual 68 Doc 065207 02 9 08...
Page 114: ...ASE 150 Operator s Manual 106 Doc 065207 02 9 08...
Page 132: ...ASE 150 Operator s Manual 124 Doc 065207 02 9 08...