20
Doc. 065207-02 9/08
2.5
Preprogrammed Methods
To help you quickly produce results with the ASE 150, Dionex provides nine
preprogrammed methods. These methods are designated by three-letter
abbreviations (see the table below).
The remainder of this section lists the recommended solvent and the operating
conditions for each preprogrammed method. You cannot edit or delete a
preprogrammed method.
Preprogrammed Method
Method Name
Semivolatiles
BNA
Total Fat (crude)
FAT
Chlorinated Herbicides
HRB
Organochlorine Pesticides
OCP
Organophosphorous Pesticides
OPP
Polychlorinated Biphenyls
PCB
Dioxins and Furans
Polymer Additives
PPE
Total Petroleum Hydrocarbons
TPH
BNA (Semivolatiles) Method Parameters
Solvent
MeCl
2
/Acetone (1:1, v/v)
Temperature
100
°
C
Static Time
5 min
Rinse Volume
60%
Purge Time
100 sec
Static Cycle
1
Cell Type
Stainless steel