ASE 150 Operator’s Manual
54
Doc. 065207-02 9/08
3.8
Performing Routine Maintenance
This section describes routine maintenance procedures that the user can perform.
All other ASE 150 maintenance procedures must be performed by qualified
Dionex personnel.
3.8.1
Daily Maintenance
•
Check the gas supply to the instrument.
•
Fill the solvent reservoir, if needed.
•
Perform a rinse cycle. After filling the solvent reservoir, always run
two rinse cycles (see
) to clear the solvent lines of air.
•
Empty the waste bottle, if needed.
•
Check for leaks.
•
Pull out the drip tray holder below the waste collection vessel
compartments (see
). If the drip tray contains liquid,
remove it from the holder, dispose of the liquid, and reinstall the tray.
3.8.2
Periodic Maintenance
NOTE The
EXTRACTION COUNTERS
screen indicates the
number of extractions the ASE 150 has performed
(see
•
Replace the PEEK seals (P/N 061687, pkg. of 50) inside the cell end
caps after approximately 50 extractions (see
•
Replace the external O-rings (P/N 049457, pkg. of 50 Teflon O-rings;
P/N 056325, pkg. of 50 Viton O-rings) in the cell end caps and rinse
cells after approximately 50 to 75 extractions (see
(Viton O-rings are for use in high temperature applications, such as
dioxin extraction.)
•
Verify that the tips of the three needles that extend from the bottom of
the needle mechanism are straight and not damaged or worn. To view
the needles, open the collection vessel compartment door and toggle
the needle switch to the
UP
Summary of Contents for ASE 150
Page 32: ...ASE 150 Operator s Manual 24 Doc 065207 02 9 08...
Page 64: ...ASE 150 Operator s Manual 56 Doc 065207 02 9 08...
Page 76: ...ASE 150 Operator s Manual 68 Doc 065207 02 9 08...
Page 114: ...ASE 150 Operator s Manual 106 Doc 065207 02 9 08...
Page 132: ...ASE 150 Operator s Manual 124 Doc 065207 02 9 08...