C • User Interface
Doc. 065207-02 9/08
129
C.1.4
Method Editor Screen
Use the
METHOD EDITOR
screen to define the parameters in a custom
method. When a preprogrammed method is selected, the method
parameters cannot be edited. All parameters do not fit on one screen. To
view additional parameters, press the down arrow button.
BYPASS HEAT-UP
Specifies whether to bypass the heat-up period for the
sample cell. The default setting (
OFF
) is recommended.
KEY SOUND
Turns on and off the beep that sounds when a keypad
button is pushed.
ERROR SOUND
Turns on and off the beep that sounds when an error
occurs.
Figure C-5. Method Editor Screen (All Parameters Shown)
Parameter
Description
METHOD
Displays the number (1 through 24) or name of the
method currently selected for editing. See
for the names and descriptions of the preprogrammed
methods.
TEMPERATURE
Specifies the temperature at which the sample cell is
heated (Off, 40 to 200 °C).
STATIC TIME
Static (extraction) time—the number of minutes (0 to
99) to maintain the cell contents (sample and solvent) at
the temperature set point.
Parameter
Description
Press the down
arrow button to
scroll through the
list of parameters.
TEMPERATURE
METHOD
STATIC TIME
60%
5MIN
100C
1
RINSE VOLUME
PURGE TIME
100SEC
1
SST
STATIC CYCLE
CELL TYPE
1
SAVE
Summary of Contents for ASE 150
Page 32: ...ASE 150 Operator s Manual 24 Doc 065207 02 9 08...
Page 64: ...ASE 150 Operator s Manual 56 Doc 065207 02 9 08...
Page 76: ...ASE 150 Operator s Manual 68 Doc 065207 02 9 08...
Page 114: ...ASE 150 Operator s Manual 106 Doc 065207 02 9 08...
Page 132: ...ASE 150 Operator s Manual 124 Doc 065207 02 9 08...