2 • Description
Doc. 065207-02 9/08
19
2.4
Method Control
A
method
determines how the ASE 150 performs the sample extraction. Methods
are defined on the
METHOD EDITOR
The following method parameters define the method. See
for
detailed descriptions of the method parameters.
•
TEMPERATURE
—the temperature at which to heat the cell.
•
STATIC TIME
—the number of minutes to maintain the cell contents (sample
and solvent) at the temperature set point.
•
RINSE VOLUME
—the amount of solvent to rinse through the sample cell after
the static extraction.
•
PURGE TIME
—the amount of time the cell is purged with nitrogen.
•
STATIC CYCLE
—the number of times the static extraction and rinse cycles are
performed.
•
CELL TYPE
—the type of cell to use (stainless steel or zirconium).
Two types of methods are available:
•
Preprogrammed
methods are application-specific methods created by Dionex.
These methods cannot be changed or deleted by the user. For a list of the
parameters for each preprogrammed method, see
.
•
Custom
methods are user-programmable methods. All custom methods
initially contain the default method parameters. You can modify these
methods as required. For instructions on how to edit a custom method, see
.
Figure 2-6. Method Editor Screen (Initial View)
TEMPERATURE
METHOD
STATIC TIME
60%
5MIN
100C
1
RINSE VOLUME
Summary of Contents for ASE 150
Page 32: ...ASE 150 Operator s Manual 24 Doc 065207 02 9 08...
Page 64: ...ASE 150 Operator s Manual 56 Doc 065207 02 9 08...
Page 76: ...ASE 150 Operator s Manual 68 Doc 065207 02 9 08...
Page 114: ...ASE 150 Operator s Manual 106 Doc 065207 02 9 08...
Page 132: ...ASE 150 Operator s Manual 124 Doc 065207 02 9 08...