
ASE 150 Operator’s Manual
100
Doc. 065207-02 9/08
5. Lift up the source needle to
remove it from the needle
block (see
then pull the needle through
the rear of the needle
assembly.
Figure 5-24. Removing the Source Needle from the Needle Block
6. Use a 1/4-inch open-end wrench to disconnect the fitting on the side of the
static valve (see
7. Remove the source needle and tubing from the ASE 150.
Figure 5-25. Disconnecting the Source Needle Fitting from the Static Valve
Source
Needle
Disconnect
this fitting
Summary of Contents for ASE 150
Page 32: ...ASE 150 Operator s Manual 24 Doc 065207 02 9 08...
Page 64: ...ASE 150 Operator s Manual 56 Doc 065207 02 9 08...
Page 76: ...ASE 150 Operator s Manual 68 Doc 065207 02 9 08...
Page 114: ...ASE 150 Operator s Manual 106 Doc 065207 02 9 08...
Page 132: ...ASE 150 Operator s Manual 124 Doc 065207 02 9 08...