17.3 IN WEIGHING
MESSAGE
DESCRIPTION
LED DISPLAY
LCD DISPLAY
WEIGHT UNDER
20 DIVISIONS
It is shown in the LCD display when one tries to print or totalise with
the 0 or negative weight.
NO ZERO CROSS OR
INSTABILITY
Printing or totalisation not possible without having the weight pass by
the net zero or by instability.
- OFF -
*** POWER OFF ***
Turning off of the instrument through the key (see section “TURNING
OFF THE INSTRUMENT”).
- OUT CAPACITY ! -
The weight is 9 divisions above the
Maximum Capacity
.
- OUT CAPACITY ! -
Approved instrument: the weight is under the gross zero (-100
divisions).
Non approved instrument: the weight is under the gross zero (-
capacity – 9 divisions).
ZEro
Execution of scale zero (see section “SCALE ZERO FUNCTIONS”).
tArE
Tare execution (see section “TARE FUNCTIONS”).
CLEAr
Tare cancellation (see section “TARE CANCELLATION”).
Prn-on
Turning on printer for maintenance (see section “TURNING ON
PRINTER IN ENERGY SAVING MODE”).
GroSS
Visualisation of the gross weight.
nEt
Visualisation of the net weight.
LoCk
Tare lock or enabled keyboard function.
unLoCk
Tare lock or disabled keyboard function.
(flashing)
Indicates that the battery is discharged (voltage at minimum level).
PRINTER ERROR:
CHECK THE CTS!
Error during the printing: totalisation not possible or print not executed
/ finished.
USER - PRESS KEY
See section “DATABASE ACCESS PASSWORD”.
.
- - - - - -
ERROR SX
CELL Y
It is shown when there is an error in the communication with the X
scale – Y cell: check the connections and communication parameters
(TECH.MAN.REF.)
.
SuM
Indicates that the weight that will be displayed is the sum weight of the
active scales.