![Dini Argeo 3590EKR Series User Manual Download Page 25](http://html.mh-extra.com/html/dini-argeo/3590ekr-series/3590ekr-series_user-manual_2505697025.webp)
3590EKR, 3590EXP, 3590EXT series indicator
E-AF08_02_13.07_EN_U
25
7.5 LOCKED/UNLOCKED/DISABLED TARE
Normally, when a tare value is inserted, this value remains stored also after the printing and resetting of the partial total
(LOCKED TARE); therefore it remains active also for the following weighs.
It is possible to choose that the tare value cancels itself automatically with each partial total printout (UNLOCKED TARE).
By pressing at length the
F5
key it is possible to quickly lock/unlock the tare:
LED DISPLAY
MEANING
LoCK
LOCKED TARE
unLoCK
UNLOCKED TARE
Furthermore it is possible to carry out this setting or disable all the tare operations, in the “TARE LOCKED/UNLOCKED”
SEtuP >>
t.LoCk
parameter of the TECHNICAL SETUP,
TECH.MAN.REF.
.
7.6 LIMITATION OF THE TARE FUNCTIONS
For specific requirements, it is possible to limit the functions of the tare
with approved instrument
; by setting “YES” in the
SEtuP
>>
d.SALE
step of the SET-UP environment
(TECH.MAN.REF.)
:
-
the SEMIAUTOMATIC TARE can not be modified with a manual or calculated tare, or one from database.
-
the manual or calculated tare, or one from database must be entered or modified only with UNLOADED scale.
With approved instrument, the
SEtuP
>>
d.SALE
step is not displayed. Read only.
7.7 TARE EXECUTION MODE OR DISABLING OF THE TARE
It’s possible to select the tare execution mode, through the setting of the
SEtuP >> t.LoCk
step
(TECH.MAN.REF.)
:
- By setting "Disable" one disables all the tare operations;
- By setting "Lock" or “Unlock”, all the tare operations previously described are enabled.
8. MULTIRANGE AND MULTIDIVISION FUNCTION
The multi range functioning
allows to subdivide the scale capacity in two or three ranges, each which is up to 3000
divisions, improving in this way the first range division in the dual range and the first two ranges in the triple range.
For example it is possible to approve the weighing system with:
-
A single range: 6 kg capacity and 2 g division (3000 div.).
-
Dual range: 6 /3 kg capacity and 2/1 g division (3000 + 3000 div.).
-
Triple range: 15 / 6 / 3 kg capacity and 5/2/1 g division (3000 + 3000 + 3000 div.).
NOTE:
For the approval of the weighing system in dual and triple ranges the cell must have better technical features in
comparison to the cell used for the approval in a single range.
This functioning is indicated by the turning on of the LED which identify the range in which one is working:
W1
first range,
W2
second range,
W3
third range (if configured); by passing to the
W2
range, the second range division is enabled; by
passing to the
W3
range, the third range division is enabled, at this point the
W1
first scale division is restored
only by
passing the gross zero of the scale
.
The multidivisional functioning
is similar to multirange, but with the difference that a range division is enabled as soon as
one enters in its range interval (in other words
without passing by the scale zero
).
NOTE:
The selection of the range number with multirange and multidivisional functioning is made during the indicator
calibration
(TECH.MAN.REF.)
.
9. DISPLAY OF METRIC DATA (inFO)
The indicator is fitted with a function named “INFO”, thanks to which it is possible to view the configuration metric data:
-
First range capacity, first range minimum weigh, first range division.
-
Second range capacity, second range minimum weigh, second range division.
-
Third range capacity, third range minimum weigh, third range division.
NOTES:
-
The minimum weigh corresponds to 20 net weight divisions.
-
The data of the second and third range appear only if actually configured.