3590EKR, 3590EXP, 3590EXT series indicator
E-AF08_02_13.07_EN_U
45
16.2.4 ALIBI MEMORY CANCELLATION (only with non approved instrument)
Command:
[II]ALDL <CR o CRLF>
in which
[II]
485 address (only in the case of 485 transmission mode)
Instrument answer:
[II]ALDLOK <CR o CRLF>
if the cancellation has been carried out effectively
[II]ALDLNO <CR o CRLF>
if the cancellation has not been carried out effectively
17. MESSAGES
While using the indicator, it is possible to incur into the following errors:
17.1 INSIDE THE SET-UP OR MENU
MESSAGE
DESCRIPTION
LED DISPLAY
LCD DISPLAY
C.Er. – 8.03
When carrying out the multirange or multidivisional calibration, the ranges have
not been entered in an increasing manner (RANGE 1 < RANGE 2 < RANGE 3).
C.Er. – 36
During the calibration some internal negative points have been calculated:
-
the calibration point is less than the zero point.
-
the signal is negative (check the connections)
C.Er. – 37
During the calibration some internal points less than the minimum value have
been calculated:
-
the calibration point is equal to the zero point.
-
a capacity too high in relation to the division has been set.
Err. – 72
Calibration error: a too low sample weight has been used.
Err. – 85
Operation not allowed because the instrument has not been calibrated.
Err.Mot
ERROR:
UNSTABLE
WEIGHT
It is shown when, during the sampling phase, the weight results to be unstable
and therefore can not be acquired.
ErPnt
ErPnt
During the acquisition of a calibration point, a null value from the converter has
been read.
SAVE ?
EXITING SETUP:
SAVE ?
The instrument requests the saving when exiting the set-up; press
ENTER
to
confirm or C to not continue.
SURE?
The instrument requests a confirmation before proceeding: press
ENTER
to
confirm or C to not continue.
Error
Generic error. Possible motives can be:
-
During the multiscale or multidivisional type of calibration, ranges in
increasing order have not been inserted (RANGE 1 < RANGE 2 < RANGE
3);
-
An error has taken place in the connection, configuration, equalisation,
calibration of the cells, etc.