PPC2+™ Operation and Maintenance Manual
Page 103
©2001 DH Instruments, Inc.
QPRR
Purpose
Read the last RPT pressure, rate. and ATM pressure.
Query
“QPRR?”
Classic “QPRR”
Remarks
The last measured ready condition, pressure rate of pressure change, and ambient
pressure for the active RPT (Lo or Hi) and range is returned in the current pressure
units. This program message is useful when a rapid response of measured pressure
is needed. Each data field is separated by a comma, and is returned in the following
order:
ready, pressure UNITS, rate UNITS/s, atm UNITS
Here are the field descriptions:
ready:
‘R’ if the current pressure ready criteria has been met,
‘NR’ if the criteria has not been met. (see the “SR” program
message)
pressure
The measured pressure for the selected RPT and range in the
current pressure units. This is followed by the current pressure units.
rate
The measured rate for the selected RPT and range in the current
pressure units per second. This is followed by the current pressure
units.
atm
The ambient pressure measured by the PPC2+ internal barometer in
the current pressure units (but always absolute). This is followed by
the current pressure units.
Example
(enhanced)
Query sent:
Query reply:
“QPRR?”
“R,2306.265 kPaa,0.011 kPa/s,97.000 kPaa”
Example
(classic)
Query sent:
Reply:
“QPRR”
“R,2306.265 kPaa,0.011 kPa/s,97.000 kPaa”
See Also
“PR”, “PRR”, “SR”
Summary of Contents for PPC2+
Page 1: ... 2001 DH Instruments Inc q 9ZO K SYX KXN 7KSX OXKXMO 7KX_KV ...
Page 14: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 8 89 9 ...
Page 24: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 8 89 9 ...
Page 84: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 76 8 89 9 ...
Page 132: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 124 8 89 9 ...
Page 150: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 8 89 9 ...
Page 156: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 148 8 89 9 ...
Page 166: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 158 8 89 9 ...