1 . 2 S P E C I F I C A T I O N S
1.2.1 GENERAL
SPECIFICATIONS
Power Requirements: 85 to 264 VAC, 50/60 Hz, 30 VA max consumption
Operating Temperature Range: 15 to 35 °C
Storage Temperature Range: -20 to 70 °C
Vibration: Meets MIL-T-28800D
Weight: 12.8 kg (28.2 lb)
Dimensions: 7.1 in. H x 12.6 in. W x 15.8 in. D (18 cm x 32 cm x 40 cm)
Microprocessor: Motorola 68302, 16 MHz
Communication Ports: RS-232 (COM1), RS-232 (COM2), IEEE-488
Pressure Ranges: Up to six pressure ranges from 0 to 5 psi (33 kPa) to 0 to 1 500 psi
(10 000 kPa) using one or two reference pressure transducers and a
barometer
Operating Medium: Any clean, dry, non-corrosive gas
Pressure Connections: Supply:
1/8 in. NPT F
Test(+):
1/4 in. NPT F
Test(-):
1/4 in. NPT F (with gauge RPT only)
Vent:
1/4 in. NPT F
Exhaust:
1/4 in. NPT F
Atm Ref:
Pass through
Pressure Limits:
Maximum Working Test Pressure:
Range H3+5 %
Maximum Test Pressure without Damage:
Range H3+10 %
Recommended Supply Pressure:
Range H3+10 %
Maximum Supply Pressure without Damage:
If Range H3>1 000 psi (7 000 kPa):
2 000 psi (14 000 kPa)
If 300 psi (2 000 kPa) <Range H3
1 000 psi (7 000 kPa):
1 400 psi (9 600 kPa)
If Range H3
300 psi (2 000 kPa):
500 psi (3 500 kPa)
1.2.2
PRESSURE MEASUREMENT SPECIFICATIONS
PPC2+ measures pressures in gauge and absolute mode in up to six ranges using one
(<3 ranges>) or two (<6 ranges>) Reference Pressure Transducers (RPTs) and a barometer.
All RPTs with ranges over 30 psi (200 kPa) are of the absolute pressure type using an
evacuated, permanently sealed reference. Absolute RPTs can measure both absolute and
gauge pressure. Gauge pressures with an absolute RPTs are defined by offsetting
atmospheric pressure with dynamic compensation for atmospheric changes using the on-
board barometer. Gauge RPTs cannot measure absolute pressure.
Summary of Contents for PPC2+
Page 1: ... 2001 DH Instruments Inc q 9ZO K SYX KXN 7KSX OXKXMO 7KX_KV ...
Page 14: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 8 89 9 ...
Page 24: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 8 89 9 ...
Page 84: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 76 8 89 9 ...
Page 132: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 124 8 89 9 ...
Page 150: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 8 89 9 ...
Page 156: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 148 8 89 9 ...
Page 166: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 158 8 89 9 ...