PPC2+™ Operation and Maintenance Manual
Page 111
©2001 DH Instruments, Inc.
UCOEF
Purpose
To convert 1 Pascal to the current pressure units.
Query
“UCOEF?”
Classic “UCOEF”
Remarks
The PPC2+ handles all pressure values internally in Pascal. The coef replied is
equivalent of 1 Pa. This program message allows the user to convert pressures
Example
(enhanced)
Query sent:
Query reply:
“UCOEF?”
“0.0010000000 kPa”
Example
(classic)
Sent:
Reply:
“UCOEF”
“0.0010000000 kPa”
See
Also Section
3.2.2
UDU
Purpose
Read or set the user defined pressure units.
Command
Query
“UDU
label, ucoef”
“UDU?”
Classic “UDU=
label, ucoef”
“UDU”
label:
User unit Device label (4 alphanumeric char maximum). It cannot be
an already supported unit.
ucoef:
“User conversion coefficient (units/Pa).
Default
“UDU USER,1.0”
Remarks
The user defined unit must be set up with the program message prior to remote or
local selection.
Example
(enhanced)
Cmd Sent:
Query reply:
“UDU MYUN, .001”
“MYUN, 0.0010”
Example
(enhanced)
Sent:
Reply:
“UDU=MYUN, .001”
“MYUN, 0.0010”
See Also
Section 3.2.2
Summary of Contents for PPC2+
Page 1: ... 2001 DH Instruments Inc q 9ZO K SYX KXN 7KSX OXKXMO 7KX_KV ...
Page 14: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 8 89 9 ...
Page 24: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 8 89 9 ...
Page 84: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 76 8 89 9 ...
Page 132: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 124 8 89 9 ...
Page 150: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 8 89 9 ...
Page 156: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 148 8 89 9 ...
Page 166: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 158 8 89 9 ...