PPC2+™ Operation and Maintenance Manual
Page 87
©2001 DH Instruments, Inc.
AINRNG
Purpose
To set or retrieve the current analog input range.
Query
Classic
“AINRNG n”
“AINRNG?”
“AINRNGn”
“AINRNG?”
Default “AINRNG
3”
Arguments
n:
The input range index
1 for 100 mv range
2 for 5 Volt range
3 for 10 Volt range
4 for 4-20 ma range
Remarks
The analog input option has 4 ranges. The selected range affects all four of the
inputs at once. When a range is changed, an auto-calibration cycle occurs, which
takes about 10 seconds.
Example
(enhanced)
Cmd sent:
Query reply:
“AINRNG 2”
“2”
Example
(classic)
Cmd Sent:
Reply:
“AINRNG=2”
“2”
Errors
ERR# 6:
The argument(s) are invalid.
See Also
Section 3.2.9
ATM
Purpose
Reads the next measured internal ambient pressure.
Query
Classic
“ATM?”
“ATM”
Remarks
The atmospheric pressure as measured by the PPC2+ on-board barometer will be
returned in the current pressure units (always absolute). This measurement is
followed by the units text.
Example
(enhanced)
Query sent:
Query reply:
“ATM?
“97.123 kPaa”
Example
(classic)
Sent:
Reply:
“ATM
“97.123 kPaa”
See
Also Section
3.4.4.7
Summary of Contents for PPC2+
Page 1: ... 2001 DH Instruments Inc q 9ZO K SYX KXN 7KSX OXKXMO 7KX_KV ...
Page 14: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 8 89 9 ...
Page 24: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 8 89 9 ...
Page 84: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 76 8 89 9 ...
Page 132: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 124 8 89 9 ...
Page 150: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 8 89 9 ...
Page 156: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 148 8 89 9 ...
Page 166: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 158 8 89 9 ...