Table 18. Troubleshooting Guide (Continued)
SYMPTOM
POSSIBLE CAUSE
SOLUTION
Front panel display is dim.
Screen saver option has
activated.
Press any key to resume full
screen power. Adjust, if
desired (Section 3.4.4.9).
Cannot access certain
functions.
>ACCESS RESTRICTED<
User levels have been set that
restrict access to certain
functions.
Change user level or consult
system manager
(Section 3.4.4.6).
Displays <FATAL ERROR> or
<FATAL FAULT>.
Encountered unresolved
internal software conflict.
Cycle power to clear. Please
record conditions leading up to
event including the numbers
displayed when [ENTER] is
pressed and report to your DHI
Authorized Service
Representative.
There is a C to the right of the
D or S control mode character
on the bottom line of the
display and it won’t go away.
The custom control function
has been used.
Reset control parameters to
default by selecting a control
mode using the CONTROL
function (Section 3.2.3).
Pressure display is flashing
and buzzer is sounding.
Current upper limit of active
range has been exceeded.
Correct overpressure condition.
Change UL and/or active range
if needed
(Sections 3.2.4, 3.2.1).
Pressure display is flashing, no
buzzer is sounding and direct
pressure control keys are
inactive.
PPC2+ has been
overpressured (pressure higher
than UL).
Correct the overpressure
condition and cycle power ON
and OFF (Section 3.2.4).
A Ready (<*>) indication is
never achieved.
Control parameter settings are
too tight and/or existing
conditions will not allow Ready
(<*>) to be achieved.
Adjust control parameters or
correct other conditions
(Sections 3.1.2.4, 3.2.6).
Will not set pressure.
Pressure and/or vacuum
supply incorrectly connected or
not adequate.
Correct pressure and/or
vacuum supply
(Sections 2.3.3, 2.3.4).
Disagreement between Hi and
Lo reference transducer or
between transducer ranges
appears excessive.
Difference is actually in
tolerance and represents
natural disagreement.
Compare differences observed
to tolerances on reference
transducer measurements
(Section 1.3.2, 3.2.6).
Summary of Contents for PPC2+
Page 1: ... 2001 DH Instruments Inc q 9ZO K SYX KXN 7KSX OXKXMO 7KX_KV ...
Page 14: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 8 89 9 ...
Page 24: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 8 89 9 ...
Page 84: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 76 8 89 9 ...
Page 132: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 124 8 89 9 ...
Page 150: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 8 89 9 ...
Page 156: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 148 8 89 9 ...
Page 166: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 158 8 89 9 ...