PPC2+™ Operation and Maintenance Manual
©2001 DH Instruments, Inc.
Page 62
Run AutoZ by Entry
AutoZ by entry allows the value of ZSTD (See Section 3.4.1, PRINCIPLE, How AutoZ
Works) to be entered manually. This provides a simple way of autozeroing relative to
an independent reference device such as a house barometer that does not interface
directly with PPC2+.
To access run AutoZ by entry press [SPECIAL] and select <1AutoZ>,
<4run>,
<1Entry>. The display is:
1. Pressure reading, units and mode of the
active RPT range.
2. Active range indicator.
3. Entry field for the value of ZSTD.
Enter the value of ZSTD in the current units.
The next display is:
The new ZOFFSET is the ZOFFSET calculated from this execution of run AutoZ.
Press [ENTER] to activate the new ZOFFSET and return to the main run screen.
Press [ESCAPE] to maintain the old ZOFFSET and return to the main run screen.
>RO`KV_OYPD900=/>S]KVaKc]NS]ZVKcONKXNOX^O\ONSX:K]MKV:K
>RO`KV_OYPD=>.S]OX^O\ONSX^ROM_\\OX^Z\O]]_\O_XS^]
2
3
ZSTD: 0.0000
New ZOFFSET: 6.5 Pa
Summary of Contents for PPC2+
Page 1: ... 2001 DH Instruments Inc q 9ZO K SYX KXN 7KSX OXKXMO 7KX_KV ...
Page 14: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 8 89 9 ...
Page 24: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 8 89 9 ...
Page 84: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 76 8 89 9 ...
Page 132: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 124 8 89 9 ...
Page 150: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 8 89 9 ...
Page 156: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 148 8 89 9 ...
Page 166: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 158 8 89 9 ...