PPC2+™ Operation and Maintenance Manual
Page 141
©2001 DH Instruments, Inc.
Hi Reference Pressure Transducer (Hi Rpt)
In single RPT PPC2+s, the RPT is designated the Hi RPT. In dual RPT models, the
Hi RPT is the RPT with the highest pressure range.
Lo Pressure Reference Transducer (Lo Rpt)
In dual RPT PPC2+s, the RPT with the lower range is designated the Lo RPT.
The Lo RPT is protected by an isolation valve and vented to atmosphere when not
in use.
5.8.1.7 PNEUMATIC
MODULE
The pneumatic module is an integrated assembly that includes two inlet (fast and
slow); two exhaust (fast and slow) control valves; vent valve and differential
pressure regulators. The differential pressure regulators use pressure feedback to
maintain a constant differential pressure across the control valves. The control valves
are solenoid type actuated by 12 V.
Figure 13. Pneumatic Module (Schematic)