PPC2+™ Operation and Maintenance Manual
Page 147
©2001 DH Instruments, Inc.
Table 18. Troubleshooting Guide (Continued)
SYMPTOM
POSSIBLE CAUSE
SOLUTION
Apparent inaccurate pressure
measurement control.
AutoZ has been run and turned
ON with an incorrect standard
for zero.
Check value of ZOFFSET.
Rerun autoZ with a valid ZSTD
reference (Section 3.4.1).
Will not vent.
System is vented but does not
indicate zero because
measurement mode is
absolute.
Check measurement mode
setting and current value of
atmospheric pressure if
absolute (Section 3.2.2)
Will not vent.
VENT port is plugged.
Open VENT port to
atmosphere.
Will not vent.
Vent valve not operating.
Have unit serviced.
Will not vent.
Purge function is activated but
no SPLT is connected or SPLT
valve is not operating.
Deactivate SPLT; correct SPLT
connection or repair SPLT
exhaust valve.
Summary of Contents for PPC2+
Page 1: ... 2001 DH Instruments Inc q 9ZO K SYX KXN 7KSX OXKXMO 7KX_KV ...
Page 14: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 8 89 9 ...
Page 24: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 8 89 9 ...
Page 84: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 76 8 89 9 ...
Page 132: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 124 8 89 9 ...
Page 150: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 8 89 9 ...
Page 156: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 148 8 89 9 ...
Page 166: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 158 8 89 9 ...