9.4.1. Sirius ACC and HS-ACC
ACC and HS-ACC modules connect to the DS-CAL test board via a BNC-to-BNC cable to the appropriate
ACC (BNC connector) channel of the DS-CAL test board.
ACC+ and HS-ACC+ modules must also have the counter connector connected to the appropriate
counter (CNT connector) channel on the DS-CAL test board.
Important
ACC+ and HS-ACC+ modules must be calibrated separately because of required different
grounding setting due to connected counter channel (see 8.5.2 Different module grounding
setting, page 42)!
Example of connecting 8xACC Sirius modules to DS-CAL test board shows Illustration 24.
Illustration 24: Example of connecting 8xACC Sirius modules to DS CAL test board
9.4.2. Sirius CHG and HS-CHG
Important
For the calibration of Sirius CHG or HS-CHG an accessory package CHG CAL ADAPTER is
required (see chapter CHG-CAL-ADAPTER)!
CHG and HS-CHG modules connect to the DS-CAL test board via a BNC-to-BNC cable to the CHG-CAL
adapter’s BNC connector. CHG-CAL adapter plugs into the appropriate MULTI (DSUB15 connector)
channel on the DS-CAL test board.