5. Connecting the calibration system together
The next step after setting up the multifunction calibrator and digital multimeter is to establish all
connections of the complete calibration system.
Caution
To avoid damage, apply at first the DC power to the DS-CAL test board and then start
connecting all the cables and doing system wiring. Disconnecting should be done in the
opposite way (unplugging the DS-CAL test board at last). It is also recommended to use a
separate power supply for the DS-CAL test board.
5.1. Digital multimeter
Connect as shown in the Illustration 6 and as described by the Table 5.
Multimeter
DS-CAL test board
Input HI
Multimeter HI
Input LO
Multimeter LO
Input I
Multimeter I
Table 5: Digital multimeter and DS-CAL test board connection