DS-CAL
SOFTWARE USER MANUAL
8. Using the DSCalibrator software
You should now see the screen as displayed in Illustration 16, showing general information about the
connected DEWESoft UUT. Here it is possible to check the firmware version, the internal temperature
and the serial numbers of the installed modules.
There are 3 possible options:
●
Adjustment
●
Calibration check
●
Adjust and Check
8.1. Adjustment
After clicking on the Adjustment button a window will appear as shown in the Illustration 17.
Here it is necessary to adjust the following settings before starting an adjustment procedure:
●
Selection of Test groups is performed automatically. Changes to the selection for an adjustment
procedure should be made only if there is a good reason to do so (see chapter Omission of tests).
●
Select the modules that you want to make an adjustment on.
●
Enter the environmental conditions (ambient temperature and humidity).
Illustration 18: Prepare check procedure window
●
Select the user (operator) that will perform an adjustment procedure (see 7.2 Users, page 32, for
adding, removing and editing users).
DS-CAL V20-1