Signal generator section:
●
Set the multifunction calibrator type to Fluke.
●
Select the correct COM port where it is connected. (only available COM ports will be shown). The
easiest way to choose the correct COM port from the USB to RS232 converter is to look for ports
that are available while the other devices are unplugged. Then start plugging in the devices one
by one and simultaneously check for available COM ports. The COM port that is new in the drop
down menu is the right one.
●
The Certificate no. and Due date can be found on the calibration certificate of your multifunction
calibrator and will be shown on the generated calibration certificates.
Multimeter section:
●
Make a selection of the digital multimeter model.
●
If you are using a 34401A digital multimeter:
●
Select the correct COM port where it is connected.
●
Enter the serial number of digital multimeter; this needs to be done manually.
●
If you are using a 34410A / 34465A digital multimeter, you need to type in the IP address that was
configured as a part of the 4.1.1 Agilent 34410A, page 20
●
The Certificate no. and Due date can be found on the calibration certificate of your digital
multimeter and will be shown on the generated calibration certificates.
Due date warning section:
Illustration 13: Database tab of the DSCalibrator
DS-CAL V20-1