DS-CAL
SOFTWARE USER MANUAL
9.3. Sirius Mini
Connect the Sirius Mini UUT as described by the Table 12. Illustration 23 shows Sirius Mini UUT as a
reference.
Sirius Mini UUT requires the the two USB cables, SYNC connector and a banana plug cable for
grounding connected.
Sirius Mini UUT
Connection
+5V Power supply
Connects to the PC.
NOTE: Only USB power is used via this connector.
USB Data
Connects to the PC.
NOTE: Only USB connection is to the Sirius Mini UUT is
established via this connector.
SYNC
Connects to the SYNC 1 connector of the DS
‑
CAL test board.
GND
Connects to the GND connector of the DS
‑
CAL test board.
Table 12: Sirius Mini UUT connection description
Important
If you are using the DS-CAL1 v1 or DS-CAL1 v3 test board, you must use and connect DS-CAN2
along with the UUT (see chapter Dewesoft UUT without CAN port)!
The next step is to connect all the modules of the Sirius Mini UUT. See chapter Sirius modules.
9.4. Sirius modules
Important
Unless otherwise noted, the channel 1 on the UUT connects to the channel 1 on the DS-CAL test
board and so on for up to all eight UUT channels, respectively.
Sirius HD UUT only: Due to doubled amount of channels two calibration passes are required. At
first pass connect UUT channels 1 to 8 to the DS-CAL test board’s channels 1 to 8, respectively.
Continue with a second pass by connecting UUT channels 9 to 16 to the DS-CAL test board’s
channels 1 to 8, respectively. You will be notified during the calibration procedure by the
DSCalibrator software which channels should be connected when.
DS-CAL V20-1
35/67