64
••••
Technical Reference section
Document code: SS1-UM-1.05
Technical Reference section
Maintenance and repair
Checking the batteries
The SunScan system requires batteries within the probe and in the Workabout. The
Beam Fraction sensor does not require batteries because it draws its power from the
probe.
The SunScan probe batteries
The probe is powered by 4 AA size alkaline cells mounted within the probe handle.
With typical use, these batteries should last from 6 - 12 months. The probe circuit
automatically "sleeps" when no readings are being taken. There is no probe on/off
switch.
Checking the probe battery level
Connect the probe to the Workabout, and from the SunData program take a reading
with the probe (any light reading will do - it is required only to activate the battery
sensing circuit).
Then press
∪
∪
∪
∪
+B
(or
Menu,Utils,About
). The screen display will show various
version numbers plus a Battery mV reading.
Replace the batteries if the reading is 4700 mV or less. A reading of 5000 mV or
above indicates that the batteries are healthy. You will get an on-screen warning if
the batteries are low. When this happens replace them as soon as you can.
If you get a 0 mV reading, the probe has not been read. Take a light reading, and try
∪
∪
∪
∪
+B
again.
Probe battery life
With a fresh set of batteries in the probe, you could take about 300,000 readings. If
you took no readings, the batteries would last for about 6-12 months.
Warning! If you are likely to store the probe for a long period, you should remove
the batteries.
Replacing the probe batteries
You must dismantle the probe handle. Disconnect the BFS and the probe's RS232
cable. The baseplate of the probe is secured to the handle by 4 cross-head corner
screws. Unscrew these to remove the handle, whilst carefully supporting the probe.
The battery holder can now be prised out of its compartment, and the batteries
changed.
The Workabout batteries
Checking the Workabout battery levels
The Workabout is supplied with a rechargeable Ni-Cd battery pack and a Lithium
back-up battery installed. Press
Shift+Ctrl+B
to display the Psion Battery info
screen. You will be warned when the batteries get low. Switch off and recharge the
batteries as soon as possible.
For full details on the care and maintenance of the Workabout's batteries you should
refer to the Workabout User Guide pages 2-5, but the following brief notes will get