OXITEST PLUS 7 OPERATING MANUAL
4.4.5
Activate Artifact Simulation
Format:
AF
n
activate artifact simulation
n
,
where
n
=
Ø for OFF
1 for movement
2 for spike
3 for tremor
Examples:
AFØ
sets the Artifact Simulation off
AF3
sets the Tremor Artifact simulation on
4.4.6
Activate Auto Preset Protocol
Format:
AP
n
Oxitest Plus 7
reverts to manual control mode with the
test protocol defined by Auto Preset protocol
n
,
where
1 <
n
<9
Examples:
AP2
sets Sp02=80%, HR=95 BPM, PA=100%
AP8
sets Sp02=97%, HR=70 BPM, PA=33%, ART= Movement
4.4.7
Download Test Report
Format:
DLTR
download test report (complete)
Example:
See example Test Report in Section 3.12.
Computer Control/Chapter 4
#
Page 33