OXITEST PLUS 7 OPERATING MANUAL
4.4
Command Definitions
For all RS-232 commands
,
be sure to follow the prescribed syntax, including leading zeros
where indicated. Failure to do so may result in unexpected operation of the
Oxitest Plus 7
.
4.4.1 Select Oximeter and Initialize Simulation Command
Format:
Ox
mmm
where
mmm
= ØØ1 for BCI 3300
ØØ2 for BCI 3300/Nellcor
ØØ3 for BCI 3302
....
Ø59 for SpaceLabs Scout with Nellcor
sensor
Example:
OXØ32
causes
Oxitest Plus 7
to select and initialize the
Nellcor N-20P patient simulation.
Note the value of
mmm
corresponding to the oximeter make/model listed above is also displayed at
the bottom right corner of the
Oxitest Plus 7
LCD when the user presses the
Pulse Ox
key on the
front panel.
The OX command will cause
Oxitest Plus 7
to select the specified oximeter whether or not the
oximeter has been ENAbled via the PULSE OX SETUP menu. The OX command also initializes
SpO
2
, heart rate and pulse amplitude to default values of 97%, 60 BPM and 100% respectively. The
status display is reset to "NO SIGNAL".
Note:
As oximeters are added to the
Oxitest Plus 7
menu by firmware upgrade, the number
mmm
may change. Be sure to check the number of any oximeter following a new
software release, and update any software programs, as necessary. The numbers given
above are for example only, and may not be representative of the current software
version.
Computer Control/Chapter 4
#
Page 31