OXITEST PLUS 7 OPERATING MANUAL
1. Specifications
1.1
Saturation (SpO
2
)
Range of adjustment is according to pulse oximeter manufacturer and/or model.
Masimo, Mindray, Nellcor, and Nellcor OEM's (e.g., Critikon; Protocol; Philips; etc.):
Variable from 35% to 100% in steps of 1%
SpO
2
presets at 35, 70, 80, 90, 93, and 97%
Selected EnviteC, Hewlett-Packard, Nihon Kohden, and Philips models:
Variable from 55% to 100% in steps of 1%
SpO
2
presets at 55, 70, 80, 90, 93, and 97%
All other manufacturers and models:
Variable from 70% to 100% in steps of 1%
SpO
2
presets at 70, 80, 90, 93, and 97%
1.2
SpO
2
Accuracy
All supported oximeters except Datex and Invivo oximeters:
55 to 100% SpO2:
±1% at specified SpO
2
presets
35% SpO2:
±2%
Nellcor accuracy specified with either DS100A sensor, or with SpO
2
calibration
selected via Nellcor Port.
Specifications/Chapter 1
#
Page 1