OXITEST PLUS 7 OPERATING MANUAL
4. Computer Control and RS-232
The operating parameters of
Oxitest Plus 7
can be controlled by an external device, through an
RS-232 serial communications port. This allows
Oxitest Plus 7
to be interfaced to an automated
tester or PC for automated oximeter testing. It also allows the control of the test parameters, data
transfer, and remote editing of Oximeter Menu, Auto Preset and Alarm Test settings.
Oxitest Plus 7
may be operated via the front panel keypad while simultaneously responding to
commands via the RS-232 port. The five control commands (Select Oximeter; Set SpO
2
; Set Heart
Rate; Set Pulse Amplitude, and Activate Artifact Simulation), and three of the download commands
(Download Manual, Auto Preset and Alarm Test Record) are all compatible with most automated
safety analyzers.
4.1
Electrical Interface
RS-232, Bidirectional,
9600
baud,
8
bits,
N
o parity,
1
stop bit.
Maximum cable length restricted to 10 feet.
Computer Control/Chapter 4
#