OXITEST PLUS 7 OPERATING MANUAL
1.
Select a Nellcor compatible pulse oximeter from the menu.
2.
Disconnect the sensor from the pulse oximeter and connect it to the Nellcor Port DB9
connector at the top of
Oxitest Plus 7
.
3.
Oxitest Plus 7
will measure the resistance in the Nellcor compatible sensor. If the
resistance is acceptable, a "sensor-detect beep" is produced and the following message is
displayed on the LCD for as long as the sensor remains connected to the Nellcor Port:
RCAL OK! Now connect
sensor to pulse ox.
4.
Disconnect the sensor from the Oxitest Plus 7’s Nellcor Port and reconnected to the
oximeter. The display returns to the Manual Mode, with the alternate "R-Curve" being
indicated by a "R" on the LCD as shown below:
N-200
R
NO SIGNAL
97% 60 BPM 100%
5.
Apply the sensor to the
Oxitest Plus 7
probe and commence testing .
If the detected resistance lies outside the acceptance range for Nellcor sensors, an "error beep" is
produced and the following error message is displayed as long as the sensor remains connected to
the Nellcor port:
ERROR: Code resistor
out of range!
In this case, the default Nellcor "R-Curve" will be selected when the sensor is removed from the
Nellcor Port.
An "error beep" is produced in response to the following error conditions:
!
detection of a resistance at the Nellcor Port when the pulse oximeter selected from the
menu is not compatible with Nellcor sensors; or
Operation/Chapter 3
#
Page 25