OXITEST PLUS 7 OPERATING MANUAL
Oxitest PLUS 7 Pulse Oximeter Tester
PULSE OX TEST REPORT
Time/Date: _________________________
Control No.: _________________________
Serial No.: _________________________
Location: _________________________
Department: _________________________
Technician: _________________________
Oxitest settings:
PULSE OX: Nellcor N-200, N-100C
SENSOR:
Nellcor W. SELECTED R-CURVE
--------Oxitest-------- --Status--
Rec. SpO
2
Rate Amp. Artf LEDs
1 97% 75BPM 100% Off OK PASS
2 93% 75BPM 100% Off OK PASS
3 90% 75BPM 100% Off OK PASS
4 97% 150BPM 30% Off OK PASS
5 97% 240BPM 15% MOVM NoRD FAIL
--------Oxitest-------- --Status--
Auto SpO
2
Rate Amp. Artf LEDs
1 97% 72BPM 100% Off OK PASS
4 97% 38BPM 100% Off OK PASS
6 97% 60BPM 0% Off NoIR FAIL
9 97% 89BPM 52% TREM OK PASS
ALARM TEST ON-TIME OFF-TIME
LOW SAT 9.8 sec 7.1 sec
HI RATE 10.6 sec 8.5 sec
LO RATE 17.4 sec 15.1 sec
LOW SIG 23.1 sec 7.4 sec
In the prior example, the results of five manual tests, four Auto Preset tests and four Alarm Tests
were saved. As shown, the
Oxitest Plus 7
signal status appears under the “LEDs” column and the
PASS or FAIL status of the test result appears at the far right. Manual test #5 shows that the
oximeter’s red LED may have been defective, which was recorded as a “FAIL”. Similarly, Auto
Preset test #6 was also recorded as a “FAIL” due to a potentially defective infrared LED in the
oximeter sensor.
If the printer is not attached, or not on-line when the
key is pressed, then the error message
shown below will be displayed. The off-line condition is tested immediately after pressing the
key and periodically while the printing is in progress.
****** ERROR ******
Printer is OFF-LINE!
The
Erase
key permits the erasure of the last test record saved (in case
PASS
or
FAIL
was pressed
at the wrong time or under the wrong conditions) or the erasure of all test records. When the
Erase
key is pressed, the display as shown below, results.
Operation/Chapter 3
#
Page 23