OXITEST PLUS 7 OPERATING MANUAL
ERASE last record?
Press YES or NO...
Pressing the
Yes
key erases the last test record, while pressing the
No
key leaves the saved
information unchanged. In both cases, the following display results.
ERASE all records?
Press YES or NO...
All test records will be erased if the
Yes
key is pressed, or no records will be erased if the
No
key is
pressed. After printing the Test Report, only the second display appears.
3.13 Using the Nellcor Sensor Port
Inside every sensor manufactured by Nellcor (Tyco) is a unique resistor which is used by Nellcor-
technology oximeters in selecting a particular SpO
2
calibration curve (R-Curve) to match the oxygen
saturation characteristic of the sensor. The Nellcor Port located at the top of the
Oxitest Plus 7
enclosure (DB9 female) allows temporary connection of Nellcor sensors to the
Oxitest Plus 7
, so
that the sensor’s calibration resistor can be measured and the SpO
2
calibration curve corresponding
to the resistor selected. Use of the Nellcor Port thus facilitates high-accuracy testing of both
disposable and re-usable Nellcor sensors, with an accuracy of ±1 to 2% across the entire range of
simulated SpO
2
levels (35% to 100%).
At power-up reset, or when a Nellcor or Nellcor OEM pulse oximeter is selected from the
Oxitest
Plus 7
Pulse Ox menu, the default Nellcor "R-Curve" is selected. After selecting the Nellcor N-
200, for example, the Manual Mode display would appear as follows on the LCD:
N-200 NO SIGNAL
97% 60 BPM 100%
As long as the STATUS display remains in the
NO SIGNAL
state, the Nellcor port is sampled by
Oxitest Plus 7
.
To use the Nellcor Port, perform the following steps:
Operation/Chapter 3
#
Page 24