OXITEST PLUS 7 OPERATING MANUAL
The effect of the
Start/Stop
key on the tester depends on the Alarm Test selected by the user, as
follows:
Prior to pushing
Start/Stop
:
Alarm Test
SpO
2
Hrt. Rate
P. Amp
Artifact
All Tests
97%
72 BPM
100%
Off
Effect of pushing
Start/Stop:
Alarm Test
SpO
2
Hrt. Rate P. Amp Artifact
#1 - LOW SAT
70%
72 BPM
100%
Off
#2 - HI RATE
97%
180 BPM
100%
Off
#3 - LO RATE
97%
40 BPM
100%
Off
#4 - LOW SIG
97%
72 BPM
3%
Off
#5 - ARTIFACT
97%
72 BPM
20%
MV
For example, the LOW SAT alarm test operates as follows:
1. The user presses the
Alarm Test
key until the LOW SAT alarm test is shown on the LCD:
LOW SAT ALM T: 00.0
97% 72 BPM 100%
2. The user starts the LOW SAT test by pressing
Start/Stop
. The alarm test immediately
switches the SpO
2
from 97% to 70%, and the timer at the top right of the LCD begins
incrementing in 0.1 second steps:
LOW SAT ALM T: 00.1
70% 72 BPM 100%
3. When the oximeter under test produces an audio alarm, the user presses the
Start/Stop
key
to stop the timer, and save the alarm ON TIME in the
Oxitest Plus 7
memory:
ALARM ON: 15.8 sec
Please wait...
Operation/Chapter 3
#
Page 20