OXITEST PLUS 7 OPERATING MANUAL
4.4.2
Set SpO
2
Command
Format:
SO
n
where
n
=
1 for 97% SpO
2
2 for 93% SpO
2
3 for 90% SpO
2
4 for 80% SpO
2
5 for 70% SpO
2
6 for 55% SpO
2
*
7 for 35% SpO
2
*
* Selected models only; see Section 1.1
Example:
SO4
sets the SpO
2
to 80%
4.4.3
Set Heart Rate Command
Format:
HR
rrr
where
rrr
=
2Ø to 25Ø BPM
Examples:
HR62
sets the Heart Rate to 62 BPM
HR175
sets the Heart Rate to 175 BPM
4.4.4
Set Pulse Amplitude Command
Format:
PA
aaa
where
aaa
=
Ø to 1ØØ per-cent
Examples:
PA7
sets the Pulse Amplitude to 7%
PA42
sets the Pulse Amplitude to 42%
PA1ØØ
sets the Pulse Amplitude to 100%
Computer Control/Chapter 4
#
Page 32