OXITEST PLUS 7 OPERATING MANUAL
Figure 12
Figure 11 - Ohmeda Flex II sensor
5) Application of Criticare Sensors
Although the Criticare 511 and 934 sensors are generally compatible with the
Oxitest Plus 7
probe (i.e. the Oxitest "finger"), the fit is less than ideal. Despite this, fairly good results can be
obtained when testing Criticare pulse oximeters by following these rules (note in all cases the
sensor LEDs are on the "bottom", and the sensor's photodiode is on "top"):
a)
Criticare 511 Sensor
An acceptable position for the 511 sensor is as follows:
Test Useage/Chapter 5
#
Page 45