Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicated theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the S/5™ CM
(1
is used exceeds the applicable RF
compliance level above, the S/5™ CM
(1
should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting of relocating the S/5™ CM
(1
.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m .
(1
S/5™ CM equipped with M-NESTPR or M-NE12STPR, M-ENTROPY, M-BIS, M-MINIC and N-DIS.
(2
For M-BIS.
Document No. M1031508
B-4(6)
Summary of Contents for S/5 Compact Anesthesia Monitor
Page 2: ......
Page 6: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 ii ...
Page 10: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 iv ...
Page 60: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 50 ...
Page 61: ...General Service Guide APPENDICES A B Document No M1031508 51 ...
Page 62: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care monitor Document No M1031508 52 ...
Page 72: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care monitor Document No M1031508 B 6 6 ...